| |
![]() | |
| | #1 (permalink) |
| Worship the feminine Current Rebreather/s: Megalodon Other Rebreather/s: Join Date: Sep 2005 Location: Den Haag (Netherlands)
Posts: 762
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | 1 point (high PO2) versus 2 point calibration Following a discussion and having learned something new, I thought I could suggest an alternative controller design that might be superior to the 2 point controllers that are out there. A single point calibration at PO2 = 1 (or thereabouts) would be superior to the current 2 point cal systems because; 1) after setting the cal, one need to only let the cell relax in air (or dil) and see that it goes back near enough to .21 (or dil O2) to know that the cell (and O2 %) is ok. 2) less prone to error. Cell non-linearity (or O2 % errors) is hidden in 2 point cals, unless one really understands the relationship between cell mV and PO2 and you do lots of data logging and calculations (or mV tables). 3) would/should be easy to design an option to in-water cal at 3-6m at your PO2 of choice. Best a value somewhere above your typical bottom PO2 The only advantage a 2-point cal has is that it slightly mitigates against cell non-linearity versus the above mentioned 1 point cal. That mitigation however is lost from the extra susceptibility to error. |
| (Offline) | |
| | #2 (permalink) |
| rEvo's daddy ![]() Current Rebreather/s: rEvo Other CCR Home Build Other Rebreather/s: rEvo Other CCR Home Build Join Date: May 2005 Location: belgium
Posts: 1,501
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: 1 point (high PO2) versus 2 point calibration just my 2c.. 2 point calibration: a solution for a problem that is not there.... 2 point calibration is needed when you have lineair behaviour and an offset cells are quite lineair between 21% and 100% O2, and have nearly no offset! so you do a 1 point calibration and check that calibration on another known point regards paul
__________________ www.rEvo-rebreathers.com .... the earth is flat, Elvis is alive, and radial scrubbers give longer dwell time than axials... |
| (Online) | |
| | #3 (permalink) |
| King of the Geeks ![]() Current Rebreather/s: MK 15.X Optima Other CCR Other Rebreather/s: Other CCR Join Date: May 2005 Location: Addison, Pennsylvania
Posts: 358
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: 1 point (high PO2) versus 2 point calibration I concur with Paul. 2 Point Cal is unnecessary. If your sensors are not linear out to 3 ATA with Zero = Zero, then you need to switch sensors or suppliers. We test the HH's to 3 ATA of O2 as part of our basic QC prior to shipping. Kevin Juergensen Juergensen Marine, Inc. |
| (Offline) | |
| | #5 (permalink) |
| Dive porn pimp ![]() ![]() ![]() Current Rebreather/s: | Re: 1 point (high PO2) versus 2 point calibration I concur with Paul. 2 Point Cal is unnecessary. What he said.If your sensors are not linear out to 3 ATA with Zero = Zero, then you need to switch sensors or suppliers. Kevin Juergensen Juergensen Marine, Inc. Brent
__________________ Attitude and self praise is no reccomendation. Dont try to be a great man, just be a man and let history be the judge of you. CHECK OUT OUR INTERWEBS FOR CUSTOM REBREATHER UPGRADES Supporting Shearwater Research Products in Europe |
| (Offline) | |
| | #6 (permalink) |
| Worship the feminine Current Rebreather/s: Megalodon Other Rebreather/s: Join Date: Sep 2005 Location: Den Haag (Netherlands)
Posts: 762
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: 1 point (high PO2) versus 2 point calibration just my 2c.. 2 point calibration: a solution for a problem that is not there.... I concur with Paul. 2 Point Cal is unnecessary. What he said. Phew, good to see that I'm not the only one who thinks soBrent .So, are you guys building systems that utilize 1 point cals? and Can you set these so you can cal at any PO2 desired (including above 1)? |
| (Offline) | |
| | #7 (permalink) |
| . ![]() Current Rebreather/s: Inspiration Classic Megalodon Classic Kiss Other Rebreather/s: Join Date: May 2006 Location: Lititz, PA
Posts: 827
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: 1 point (high PO2) versus 2 point calibration Phew, good to see that I'm not the only one who thinks so Gilles .So, are you guys building systems that utilize 1 point cals? and Can you set these so you can cal at any PO2 desired (including above 1)? The HH (used on the Optima, HH CCR and as a add-on) is one point and also the Inspiration Classic. I suspect the other AP CCR's but I do not own them so I am not sure. John |
| (Offline) | |
| | #8 (permalink) |
| flap-flop ..... flap-flop Current Rebreather/s: rEvo Other Rebreather/s: rEvo Join Date: Nov 2005 Location: Denmark
Posts: 406
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: 1 point (high PO2) versus 2 point calibration So, are you guys building systems that utilize 1 point cals? The rEvo-dreams only calibrates at 100%. and Can you set these so you can cal at any PO2 desired (including above 1)? Allthough not specifically stated in the manual, I allways calibrate / check calibration in 100% during assembly, and when assembly is done, I remove exhale-hose (P-connector - very easy), and breathe air into loop. Assembly is then done when all 4 cells steadily reads 0.21 -> Ie. I verified that my calibration makes sense. ![]() This allso ensures auto-shutdown of the rEvo-dreams, after 15mins, and generally is better than leaving the system at 100% wich will shorten the cell-life. Now if I were to use say a 95% gas, I would do the same, but exspect it to read 0.208*100/95 = 0.22. While diving I'd set the deco to run at a factor 95/100 lower PPO then read on the displays. No biggie. Nicolai
__________________ Woohooo - I can change my rEvo!Its going to be bitchin' tricked out piece of gear |
| (Offline) | |
| | #9 (permalink) |
| Worship the feminine Current Rebreather/s: Megalodon Other Rebreather/s: Join Date: Sep 2005 Location: Den Haag (Netherlands)
Posts: 762
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: 1 point (high PO2) versus 2 point calibration Kinda boring when everyone is in agreement .Perhaps I can stir some fury in the Meg forum .I would like a single point cal system, that has options allowing in water PO2 cals up to 1.6 ATA. I have a 2-point system and I find that it sucks. Have to be meticulous with the mV readings to check cells. Dive safe ![]() |
| (Offline) | |
| | #10 (permalink) |
| flap-flop ..... flap-flop Current Rebreather/s: rEvo Other Rebreather/s: rEvo Join Date: Nov 2005 Location: Denmark
Posts: 406
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: 1 point (high PO2) versus 2 point calibration I would like a single point cal system, that has options allowing in water PO2 cals up to 1.6 ATA. Why?At surface you have near constants pressure. You dont have your lungs on/offgassing into the loop. It would make no sense doing something as complicated as calibrating (with precision), in water, on existing rebreathers. Only when designed for it it "perhaps" makes sense. Ie. the 'POS Mk VI', where the loop is not flushed only the cell-chamber. If calibrated at 100%@surface youre calibrating at 77% of normal scale (1.3). At the beginning of the dive you have the option to do yet another verification-point, diving pure O2 at 6m. before descending - I know zerogravity diving is advocating this method. To summarize:
![]() Still cant see why you want calibration during the dive - if so, then sign up for a PoS Mark VI ![]() Regards Nicolai
__________________ Woohooo - I can change my rEvo!Its going to be bitchin' tricked out piece of gear Last edited by Hanssing : 26th January 2008 at 11:45. |
| (Offline) | |