| | #3 (permalink) |
| Worship the feminine Current Rebreather/s: Megalodon Other Rebreather/s: Join Date: Sep 2005 Location: Den Haag (Netherlands)
Posts: 762
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: MV Specs Does anyone have the tolerances in MV for air and oxygen? If your cells are linear, the O2 output should be 0.99/0.209 x the mV output in air.Thanks Marty For e.g. Air = 9.0 mV, O2 purity = 98% O2 mV = 0.98/0.209 x 9.0 mV = 42.20 mV |
| (Offline) | |
| | #4 (permalink) |
| O2ptima Test Dummy ![]() ![]() Current Rebreather/s: Inspiration Classic Megalodon Sport Kiss Classic Kiss Optima Titan Other Rebreather/s: Join Date: Sep 2005 Location: Florida
Posts: 367
![]() ![]() ![]() ![]() ![]() ![]() | Re: MV Specs 8 to 12 MV air if its less than 8 replace 40 MV min on O2 if less it will not calibrate |
| (Offline) | |
| | #5 (permalink) |
| CCR Instructor Current Rebreather/s: Inspiration Classic Optima Other Rebreather/s: Inspiration Vision Evolution Join Date: Apr 2005 Location: FL, USA
Posts: 191
![]() | Re: MV Specs 8 to 12 MV air if its less than 8 replace Just had 2 new sensors in package fail calibration, ran at 30mv and 36mv in O2. People at AI Inc are great in cross shipping replacements.40 MV min on O2 if less it will not calibrate |
| (Offline) | |