It appears you have not yet registered with our community. To register for free click here
Rebreather World
       
Go Back Rebreather World Rebreather Models CCR - Closed Circuit Rebreathers Innerspace Systems Corp. Megalodon Rebreather

Megalodon Calibration Procedures



Reply
 
LinkBack Thread Tools Display Modes
Old 6th April 2006, 16:29   #41 (permalink)
RBW Founder
 
schford's Avatar

Current Rebreather/s:
Megalodon
Sport Kiss

Other Rebreather/s:
 
Join Date: Jan 2005
Location: UK
Posts: 3,564
schford has a reputation beyond repute schford has a reputation beyond repute schford has a reputation beyond repute schford has a reputation beyond repute schford has a reputation beyond repute schford has a reputation beyond repute schford has a reputation beyond repute schford has a reputation beyond repute schford has a reputation beyond repute schford has a reputation beyond repute schford has a reputation beyond repute
Megalodon Calibration Procedures

Quote: (Originally Posted by Drmike)
Do all Meg users use the same calibration method/kit?
APECS and Shearwater differ in so much as Shearwater is one point calibration (Done in 100% O2) and APECS is 2 point done in air and 100% o2 on both handsets I believe....

The kit is an optional extra that was only brought out about a year ago - before that it was done by trying to bring the loop to 100% o2

In terms of calibration though - if I have thought this through correctly, if you mess up the calibration on the (shearwater) Meg then I think that your displayed PPo2 will be higher than it is in the loop - so am not sure how you would get o2 convultions (at least on the Shearwater with 1 point cal - I have to admit am not sure how 2 point cal woudl affect it)...

eg when you calibrate if the flow of gas is restricted (eg by an anyliser) then you would actually be under a slight pressure in the head so when you set it at .98 or whatever it was it could actually be less eg .9 at ambiant - this could affect your deco but should not cause convultions from breathing to high a PPO2 ...
__________________
Bailout and Kit Management account for Murphy's Law

The only bad question is one you did not ask and later wish you had.

Use of RBW is subject to the Rebreather World Terms & Conditions of Use

Last edited by ROB DAVIE : 7th April 2006 at 03:56.
(Offline)
 
Digg this Post!Add Post to del.icio.us
Reply With Quote
Old 6th April 2006, 16:29   #42 (permalink)
RBW Member
 
Mark Chase's Avatar

Current Rebreather/s:
Classic Kiss

Other Rebreather/s:
Inspiration Classic
 
Join Date: Jan 2005
Location: Kent
Posts: 2,858
Mark Chase has a reputation beyond repute Mark Chase has a reputation beyond repute Mark Chase has a reputation beyond repute Mark Chase has a reputation beyond repute Mark Chase has a reputation beyond repute Mark Chase has a reputation beyond repute Mark Chase has a reputation beyond repute Mark Chase has a reputation beyond repute Mark Chase has a reputation beyond repute Mark Chase has a reputation beyond repute Mark Chase has a reputation beyond repute
Megalodon Calibration Procedures

Quote: (Originally Posted by deactivating)
Seems to me that there are a lot of rebreather divers out their that do not fully understand how to calibrate their units. I have come across this problem with certified inspiration divers who do not know exactly what they are doing during calibration, only selecting 98% every time, not analyzing their tanks not adjusting for altitude (classic handsets) and not understanding what is happening as the unit begins calibration just staring at the handsets with no real idea of what’s happening.

Whats to know?

Do you wish to calibrate?

Yes

Open Mouthpiece?

Yes

Open 02?

Yes

The hand sets ask you for the milibar setting on the start up routine and you put in 1020 if you don't have a VR3 handy and yes we always put in 99% 02 because 02 is rarely 100% at sea level and it alow for a bit of contamination in the head. Your on the boat about to dive so your normally about 1.5m above sea level

But how much does any of this relay matter?

Calibrating on 98% instead of 100% will give you a slightly rich mix by a minuscule 2%. It would be amazing to me if the unit was accurate within 2% any way. I would be surprised if an error of 0.05pp02 made much if any difference in the scheme of things.

I am not sure if refresher training will help divers who couldn't understand how to calibrate a unit despite passing Mod1 and Mod3?

ATB

Mark Chase
__________________
See my "Doing It Chasey" video where I'm locked into a padded room, naked, with two ball bearings and within an Hour, I manage to lose one and break the other!!!

Kevin Juergensen 16/11/08
[/quote]

Last edited by ROB DAVIE : 7th April 2006 at 03:57.
(Offline)
 
Digg this Post!Add Post to del.icio.us
Reply With Quote
Old 6th April 2006, 16:39   #43 (permalink)
Blogs Admin / Forum Mod
 
sven's Avatar

Current Rebreather/s:
Inspiration Classic

Other Rebreather/s:
Evolution
 
Join Date: Nov 2005
Location: Sydney
Posts: 551
sven is a glorious beacon of light sven is a glorious beacon of light sven is a glorious beacon of light sven is a glorious beacon of light sven is a glorious beacon of light sven is a glorious beacon of light sven is a glorious beacon of light sven is a glorious beacon of light sven is a glorious beacon of light sven is a glorious beacon of light sven is a glorious beacon of light
Send a message via MSN to sven Send a message via Skype™ to sven
Megalodon Calibration Procedures

Quote: (Originally Posted by tecdivertraining)
Seems to me that there are a lot of rebreather divers out their that do not fully understand how to calibrate their units. I have come across this problem with certified inspiration divers who do not know exactly what they are doing during calibration, only selecting 98% every time, not analyzing their tanks not adjusting for altitude (classic handsets) and not understanding what is happening as the unit begins calibration just staring at the handsets with no real idea of what’s happening.

Clearly understanding how to calibrate the unit is essential. This skill needs to be addressed a lot more during entry-level CCR courses.

Maybe certified divers should be asked to attend some kind of update periodically to learn more about the developing world of diving CCR and gain up to date information on the units they dive.

For some RBW is a great tool for learning but not all rebreather divers visit this great forum to continue their education, in my opinion divers skills should be periodically assessed and new standards and procedures reviewed ensuring divers stay up to date with the evolving world of Rebreathers.

Just my two pence worth
Matthew, Joe.

In principle, I agree with what you have said. Unfortunately, the Internet is frequented by "Internet divers", who don't take proper training and believe, that they know it all after reading a few pages on the Net ... I think it would be better to influence certification agencies to improve their training, rather than to publish this information here. If this site was only for certified divers who want to exchange information and improve their knowledge ... different story.

Let the flame wars begin
__________________
Regards,
Sven

The Sydney Project website: http://www.sydneyproject.com
My Blog: http://sven.rebreatherworld.com

"Whenever you find yourself on the side of the majority, it's time to pause and reflect." Mark Twain

Last edited by ROB DAVIE : 7th April 2006 at 03:57.
(Offline)
 
Digg this Post!Add Post to del.icio.us
Reply With Quote
Old 6th April 2006, 17:03   #44 (permalink)
Pacific Northwest
 
UWSojourner's Avatar

Current Rebreather/s:
Megalodon

Other Rebreather/s:
 
Join Date: Feb 2005
Location: Portland Oregon
Posts: 558
UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future
Megalodon Calibration Procedures

Quote: (Originally Posted by jradomski)
I see several potential problems with calibrating the meg.. I verified several of these while I was at the factory..
Withdrawn comment.

Last edited by UWSojourner : 13th May 2006 at 01:00. Reason: No longer believe back-pressure a significant problem using the cal kit of the Meg
(Offline)
 
Digg this Post!Add Post to del.icio.us
Reply With Quote
Old 6th April 2006, 17:13   #45 (permalink)
RBW Member
 
teomannaskali's Avatar

Current Rebreather/s:
Home Build

Other Rebreather/s:
Not Bought Yet
Home Build
 
Join Date: Feb 2006
Location: Istanbul TR
Posts: 421
teomannaskali is just really nice teomannaskali is just really nice teomannaskali is just really nice teomannaskali is just really nice teomannaskali is just really nice teomannaskali is just really nice teomannaskali is just really nice teomannaskali is just really nice teomannaskali is just really nice
Send a message via MSN to teomannaskali
Megalodon Calibration Procedures

Quote: (Originally Posted by Genesis)
Second, I don't like two-point calibration. I know why some people DO like it, but the engineer in me doesn't like it. My reason is that the cell is supposed to have a 0mv output in 0% O2. That is, the slope of the response line is set by a single-point calibration but one end-point is fixed.

Two-point calibration does not fix either end point, but rather defines the slope with the two points of calibration and allows the end point to float. Unless the software verifies that the "zero" end point is really zero, there is a danger here that you could easily set a slope that does not conform to reality. It would be relatively easy to end up with a significant error, especially in the case where one or more cells are malfunctioning, or there is a problem with the signal conditioning circuit - yet the "calibrate" can be performed. Looking at the mv levels does not address this, as I've got three cells here which are all "good" and yet which have significant variability in mv output levels; the "range" of acceptable mv levels for functional cells is in the 25% area! That of course would be a HUGE error if that ended up in the displayed (and controlled) value.....

The data sheet of the O2 sensors doesnt say anything about 2 point calibration.

It states that if you calibrate it with air 20.9 you should get accurate readings up oh 1 ata. And if you calibrate it with %100 o2 you will get accurate readings up to 2 ata. (Check the datasheets to see how accurate accurate is)

[quote=Genesis]
First, O2 fuel cells are QUITE sensitive to humidity. You want to see this? Its quite easy. Take a tank full of AIR (which is very dry, with RHs in the single digits if the dewpoint is at -50F or better where it should be.) Calibrate your O2 analyzer on atmospheric air to 20.9, then plug in the tank air and analyze it. Notice anything? Heh, how come it doesn't read 20.9 any more? You have two samples, both taken from the same air. The only difference is that one's dry, the other's moist. Uh huh....

This is especially bad with a Rebreather because the loop is almost always at or very near 100% RH. This is an immediate source of error and there's simply no way around it, because your source gas for calibration on a "known source", if using some form of tank gas (e.g. O2) is going to be dry while the ambient environment moist - and humidity makes the cells read LOW.
[quote]

The data sheet claims that the sensors are not affected by non-condensing humidity. If it condenses thet the O2 cant be absorbed by the sensor and result in a low reading.

The cell accurately measures the ppO2 of Oxygen (again see the data sheet for how accurate). What happens is (I think, I dont claim to know it all) when you use dry air you have 20.9 O2 in the mix and 79.1 N2. But when you do it in humid air you are adding water vapour to what you are measuring. Since the total pressure is the same (1 atm at sea level) this means that there now 3 gasses in where there was 2. So the partial pressure of O2 drops because there really is less of 02. So the sensor measures correctly.

So in a rebreather (on the surface) if there is humidity in the loop there is no way you can get a %100 O2 level.

Teoman

Last edited by ROB DAVIE : 7th April 2006 at 03:58.
(Offline)
 
Digg this Post!Add Post to del.icio.us
Reply With Quote
Old 6th April 2006, 17:20   #46 (permalink)
Life starts at 100
 
rogeringebo's Avatar

Current Rebreather/s:
Pelagian

Other Rebreather/s:
Megalodon
Classic Kiss
Pelagian
 
Join Date: Sep 2005
Location: Stockholm Sweden
Posts: 739
rogeringebo is just really nice rogeringebo is just really nice rogeringebo is just really nice rogeringebo is just really nice rogeringebo is just really nice rogeringebo is just really nice rogeringebo is just really nice rogeringebo is just really nice rogeringebo is just really nice rogeringebo is just really nice
Megalodon Calibration Procedures

When I dived my KISS I always filled the loop up with 100% O2 and went down on the bailoutreg to about 6m and made sure that the sensors could come up to 1,6 or close to that....then dil flush and then continue the dive..quite a good way to make sure the sensors can get up to that number..(normally when getting old that would be a problem)
__________________
Life starts at 100

Last edited by ROB DAVIE : 7th April 2006 at 03:58.
(Online)
 
Digg this Post!Add Post to del.icio.us
Reply With Quote
Old 6th April 2006, 17:28   #47 (permalink)
New Member
 
hyrax's Avatar

Current Rebreather/s:
Not Bought Yet

Other Rebreather/s:
Not Bought Yet
 
Join Date: Dec 2005
Location: NC. USA
Posts: 21
hyrax is an unknown quantity at this point
Megalodon Calibration Procedures

First let me say I’m not YET a rebreather diver and I have not YET built my own loop controller for a rebreather. I am however a professional programmer. I think that it might help if everyone understands what you are doing when you calibrate a PPO2 sensor.

First this is what we are solving, the equation for the line the sensor would make if you graphed the PPO2 on the y axis and the mV reading on the x axis.

y= mx+b

y is your PPO2
m is your mV reading
x is the slope of your line
b is your offset

For a 2 point calibration.

Given on air your PPO2 should be about .21. Lets just say our mV reading is 20. For our O2 to make it simple lets use 1.0 for our PPO2 and 50 for our mV reading. First we want to find x that is just our slope between our first and second point. Remember that grade school math slope is rise over run. In our case.

x = (1.0 - .21)/(50 – 20)

and we get

x = 0.0263333

Now we have to solve for b. So let’s use our numbers we used for the air calibration.

.21 = 20 * 0.0263333 + b

b = .21 – (20 * 0.0263333)

b = -0.316666

So we get:

y = 0.0263333 * x – 0.316666

So 40 mV looks like this.

Y = 0.0263333 * 40 – 0.316666 or PPO2 = 0.736666

If I’ve made any mistakes feel free to tell me what an idiot I am, I have thick skin I can take it.

Last edited by ROB DAVIE : 7th April 2006 at 03:59.
(Offline)
 
Digg this Post!Add Post to del.icio.us
Reply With Quote
Old 6th April 2006, 17:34   #48 (permalink)
Pacific Northwest
 
UWSojourner's Avatar

Current Rebreather/s:
Megalodon

Other Rebreather/s:
 
Join Date: Feb 2005
Location: Portland Oregon
Posts: 558
UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future UWSojourner has a brilliant future
Megalodon Calibration Procedures

Quote: (Originally Posted by Genesis)
First, O2 fuel cells are QUITE sensitive to humidity. You want to see this? Its quite easy. Take a tank full of AIR (which is very dry, with RHs in the single digits if the dewpoint is at -50F or better where it should be.) Calibrate your O2 analyzer on atmospheric air to 20.9, then plug in the tank air and analyze it. Notice anything? Heh, how come it doesn't read 20.9 any more? You have two samples, both taken from the same air. The only difference is that one's dry, the other's moist. Uh huh....
Shouldn't it be that way?

If water vapor makes up part of the atmospheric pie we breath, then don't you want the analyser to read lower even if 100% O2 is the source gas? For example, if after calibrating to dry 100%O2 you take a reading in high humidity with O2 the source gas I would expect a reading of about 95% I think.

If I'm thinking about this correctly I think you want the dry calibration AND you want the reading to drop with introduction of humidity. If so, the low end of the two point calibration should also occur with dry gas from an take filled with AIR.

Last edited by ROB DAVIE : 7th April 2006 at 03:59.
(Offline)
 
Digg this Post!Add Post to del.icio.us
Reply With Quote
Old 6th April 2006, 17:56   #49 (permalink)
Bubbless Box of Death
 
Genesis's Avatar

Current Rebreather/s:
Home Build

Other Rebreather/s:
Home Build
 
Join Date: Oct 2005
Location: Sunny Florida
Posts: 1,453
Genesis is a splendid one to behold Genesis is a splendid one to behold Genesis is a splendid one to behold Genesis is a splendid one to behold Genesis is a splendid one to behold Genesis is a splendid one to behold Genesis is a splendid one to behold Genesis is a splendid one to behold Genesis is a splendid one to behold Genesis is a splendid one to behold Genesis is a splendid one to behold
Megalodon Calibration Procedures

Quote: (Originally Posted by UWSojourner)
Shouldn't it be that way?

If water vapor makes up part of the atmospheric pie we breath, then don't you want the analyser to read lower even if 100% O2 is the source gas? For example, if after calibrating to dry 100%O2 you take a reading in high humidity with O2 the source gas I would expect a reading of about 95% I think.

If I'm thinking about this correctly I think you want the dry calibration AND you want the reading to drop with introduction of humidity. If so, the two point calibration should also occur with dry gas from an take filled with AIR.
Exactly.

My point is that for two-point to be accurate you need the cal to take place with gases that have the same RH! Otherwise you're setting up an offset error - the slope is now wrong because you've introduced an offset into the calibration that is fictional (the cell must produce 0mv in 0% O2, as its an electrochemical cell that relies on the O2 for its activity)

If you read 20.9 in "atmospheric" air but its really 20.6 (due to the water vapor) then you've depressed the slope. If you cal with dry O2 from a tank for the other side you've now offset the "zero point" high, which causes the unit to read below actual. You've pushed both ends of the slope in opposite directions, which makes the errors additive. In effect, the two-point cal becomes less accurate than a single-point because the errors in gas use are additive to the error in the result.

Now that alone shouldn't be enough to tox you - but if you now have positive pressure in the sensor areas, you now are forcing the slope even lower; the errors are again additive. With a 0.5psi positive pressure in the head you'd have introduced about a 3.5% error in the reading. Add to that another couple of percent for the RH offset (ambient air cal on a humid day in Florida) and you could easily be off by 5% or so - close to a full tenth off at a 1.6 PO2.

This STILL shouldn't be enough to "get you". But - if you combine this with a non-linear response to hyperbaric PO2s due to cell fault or age......

If you studiously validate the hyperbaric PO2 readings before relying on them then you're PROBABLY ok, as you'll catch any gross errors and smaller ones aren't material enough to matter. But if you don't........

I put two-point cal into my controller's firmware and found that I didn't like the offset values that resulted at the zero end. The algabraic equation should be X = c(PO2); any "offset" is by defintion an error.

Reducing it to single-point resulted in an error of less than 2% at 1.6 PO2, and within the 1% tolerance of the sensor at a 1.0 PO2, all calibrating using dry (tank) air. Using atmospheric air results in an error of less than 3% full scale, which is less than half of one significant digit (that is, it won't result in a 1.6 actual PO2 being 1.5 nor 1.7). Two-point calibration frequently DID lead to errors large enough to affect the display, especially when tank O2 was used for the high point and humid atmospheric air for the low. This makes sense, because the errors are additive.

All these cals are currently being done with the sensors "loose" as the head's not yet done, so there's no possible positive-pressure issue to deal with (yet)

If you do two-point calibration and wind up with an equation that is X = c(PO2) + b, for any value of "b" that is material you need to fail the cal, because that equation is known false. The slope must intercept 0,0 for the calibration to be valid.
__________________
"A venturesome minority will always be eager to get off on their own, and no obstacles should be placed in their path; let them take risks for Godsake, let them get lost, sunburnt, stranded, drowned, eaten by bears, buried alive under avalanches - that is the right and privilege of any free American."
http://www.denninger.net
http://www.diversunion.org/liability.htm - Fix the Diving Cert racket

Last edited by ROB DAVIE : 7th April 2006 at 04:00.
(Offline)
 
Digg this Post!Add Post to del.icio.us
Reply With Quote
Old 6th April 2006, 18:37   #50 (permalink)
Third Class Snorkeller
 
DonalWalsh's Avatar

Current Rebreather/s:
Megalodon

Other Rebreather/s:
Evolution
 
Join Date: Nov 2005
Location: Moving to London in Sept.....which way to the South Coast??
Posts: 133
DonalWalsh has a spectacular aura about DonalWalsh has a spectacular aura about DonalWalsh has a spectacular aura about DonalWalsh has a spectacular aura about DonalWalsh has a spectacular aura about DonalWalsh has a spectacular aura about
Megalodon Calibration Procedures

Quote: (Originally Posted by jradomski)
no connect the analyzer to monitor the progress, once it reaches the proper reading DISCONNECT it then calibrate.
Gotcha- that makes more sense.

Last edited by ROB DAVIE : 7th April 2006 at 04:00.
(Offline)
 
Digg this Post!Add Post to del.icio.us
Reply With Quote
Reply


Thread Tools
Display Modes

Posting Rules
You may not post new threads
You may not post replies
You may not post attachments
You may not edit your posts

vB code is On
Smilies are On
[IMG] code is On
HTML code is Off
Trackbacks are On
Pingbacks are On
Refbacks are On



RebreatherWorld.Com ©2005 - 2008
Rebreather World, RBW and the Rebreather World Logo are Trademarks
All rights reserved, no republishing of content without written permission.
By using this website you have agreed to our Terms & Conditions of Use

Search Engine Optimization by vBSEO 3.1.0