| |
![]() | |
| | #141 (permalink) |
| flap-flop ..... flap-flop Current Rebreather/s: rEvo Other Rebreather/s: rEvo Join Date: Nov 2005 Location: Denmark
Posts: 408
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: ISC posts test data Hi Alex, As allways it is a pleasure to read your informed posts. I have several times wondered about what you write below, wich is completely in line with what you have written before regarding WOB and scrubber-usage: 1. State your test conditions clearly. From the ISC figures it is clear they used fresh lime: the WOB tests should have been done with near exhausted lime because WOB risks from CO2 retention + O2 risks are most acute when the lime is nearing exhaustion. There is a big difference in the figures between fresh lime and exhausted lime in most scrubbers. This is important: with near expired lime our tests show it does not quite meet EN14143 WOB limits, but here with new lime, ISC are showing it does meet EN14143 WOB limits. Similarly, the attitude of the mannequin is not stated. One of the keypoins we keep returning to is comparable test-results based on comparable test-conditions! I therefor have a hard time understanding if it is your oppinion or the ENxxxx specifications that dictates the WOB-messuremeants with used lime?The reason is, I see "near exhausted lime" as hard to define and reproduce for a testing-house. In essence you only have one go to perform the meassurement - right a breakthrough. IMO, it would make more sense that the standard used an clearly defined scubber-state wich is offcourse unspent, and then set the pass-levels so they would encompass the expected deterioration in WOB through scrubber-life. From a safety-perspective i completely agree with the notion that things should be tested in worst-case scenarious, but from a meassuring perspective this gives a "wobbly" foundation for the WOB-figures.... Even the packaging density is probably hard enough to get comparable for user-packed scrubbers, that comparing result makes sense at all. But ofcourse the packaging should be the same for all tests, wob and duration etc.... BTW: It is clear that testing with fresh lime would rob the EAC-designs (like the OR) of one of its advantages; that the deterioration with use is much less that granular. So my questions boils down to:
Nicolai Hanssing
__________________ Woohooo - I can change my rEvo!Its going to be bitchin' tricked out piece of gear |
| (Offline) | |
| | #142 (permalink) |
| So much more to learn ![]() Current Rebreather/s: | Re: ISC posts test data One of the keypoins we keep returning to is comparable test-results based on comparable test-conditions! I therefor have a hard time understanding if it is your oppinion or the ENxxxx specifications that dictates the WOB-messuremeants with used lime? The standards require testing in the worst case normal operating conditions. They do not define the details in every case.The reason is, I see "near exhausted lime" as hard to define and reproduce for a testing-house. In essence you only have one go to perform the meassurement - right a breakthrough. IMO, it would make more sense that the standard used an clearly defined scubber-state wich is offcourse unspent, and then set the pass-levels so they would encompass the expected deterioration in WOB through scrubber-life. From a safety-perspective i completely agree with the notion that things should be tested in worst-case scenarious, but from a meassuring perspective this gives a "wobbly" foundation for the WOB-figures.... Even the packaging density is probably hard enough to get comparable for user-packed scrubbers, that comparing result makes sense at all. But ofcourse the packaging should be the same for all tests, wob and duration etc.... BTW: It is clear that testing with fresh lime would rob the EAC-designs (like the OR) of one of its advantages; that the deterioration with use is much less that granular. So my questions boils down to:
Nicolai Hanssing The standards have WOB tests separate from scrubber life tests. They are not specific enough in this area. This means that people can bypass the intent, which is to test under worst case normal operating conditions, but using fresh lime. The WOB tests do not involve CO2 generation, but are at 4C (this makes no difference to WOB in my experience). Alex |
| (Offline) | |
| | #143 (permalink) |
| So much more to learn ![]() Current Rebreather/s: | Re: ISC posts test data Here is the data I promised to post. ISC published mouthpiece figures, but the conditions were not too clear. I understand Golem Gear and others have an interest in this detail, so here are our independent mouthpiece figures for Meg, Inspo original, Dolphin and DL mouthpieces. The Ouroboris should be very similar to the DL figure (neck and neck on that component, as the geometry is almost the same and both use very floppy mushrooms). Sorry I do not have the Golem BOV here to compare with. The mouthpiece test is the EN14143:2003 test, applied to just the mouthpiece with its mushroom valves fitted, but no hoses other than from the diver's mouth to the breathing emulator. This measures the worst case WOB at 40msw using air, at an RMV of 75 litres per minute with a tidal volume of 3 litres. We also include the EN14143:2003 calibration data (at 50msw, 62.5 litres per minute, as per the calibration point stated in EN14143). For summary, the figures are:
This includes some WOB from the breathing emulator itself. It emulates the work of breathing inside the diver. This should be included: with the calibration orifice the total WOB is measured at exactly 3.3 J/L. If you want to consider the device under test, without the work of the diver, then deduct 0.6 J/L at 40msw, or 0.72 J/L at 50msw. A final caution: though the resistance that causes WOB from the breathing emulator and device under test are serial, they do not necessarily add up linearly (the breathing emulator may contribute less out of the total, because WOB figures are the integral within the Lissajou). The Lissajou figures we publish shown go the normal way around (i.e. as in NORSOK, STANAG, NEDU etc), opposite direction to EN14143. If you want the EN14143 image, mirror it (show it in a mirror). Hope this is helpful in demonstrating how to do a direct comparison of different units, and present the results in a way that anyone can check for themselves if they have access to the appropriate equipment. The mouthpiece, counterlungs, scrubber, hoses and connectors are the parts that determine the total WOB of a rebreather. We provide a full design verification report on the breathing emulator. You can see from the accuracy in which it tackles the EN14143:2003 orifice, it is pretty well spot on. It is a very good thing for companies to test their equipment and publish the data. By doing this openly, we can keep each other honest: mistakes are easily made. Just to be clear, I asked for all these tests to be repeated over the past couple of days, and I compared the results with those we had obtained much earlier: they are the same. I personally witnessed and checked the test conditions, in addition to those named on the certificates. Thanks to Bob Davidov and Misha Soloviev for the drudgery of repeating these tests just so I could be double sure. Alex Last edited by AD_ward9 : 3rd April 2008 at 14:49. |
| (Offline) | |
| | #144 (permalink) |
| Who loves ya, baby ![]() ![]() Current Rebreather/s: | Re: DL posts test data Thanks for the breathing figures, Alex, much appreciated.
__________________ Cheers Stefan "Political Correctness is a doctrine, fostered by a delusional, illogical minority, and rabidly promoted by an unscrupulous mainstream media, which holds forth the proposition that it is entirely possible to pick up a turd by the clean end.!" |
| (Offline) | |
| | #145 (permalink) |
| So much more to learn ![]() Current Rebreather/s: | Re: ISC posts test date In reference to rEvo: 3.15 J/L is just within EN14143, but it is still pretty high. Paul, I need to correct myself. I misread your test conditions (when I wrote the above, I had them in my mind from another thread that the test pressure was barg but you state here it is bar absolute, at the correct pressure of 5 bar: 40msw). At 3.15J/L at 5 bar absolute (approx 40msw) and 75l/min RMV, the rEvo fails to meet limit for WOB in EN14143:2003 - the limit under these conditions is, as you know, 2.75 J/L. It would also fail all other major standards and guidelines: NORSOK U101, STANAG 1410 with the 2007 NEDU limits, US Navy 1994 limits etc. I understand your result is for the diver vertical position, which should be the worse of the two test orientations on the rEvo if the test has been carried out properly. However, if it helps, all the other sports rebreathers also fail, except the Ouroboros, Sentinel and the Open Revolution rebreathers (both dual and single scrubbers). It would seem only fair for CCR Ltd to update their comparison with your figure, and also put you in the right place of the table: at the moment even with your previous figure of just below 4J/L, your position in the CCR Ltd table was two columns to the left of where you should have been. Now you should fit between the Inspo and the Evo. CCR Ltd's table is below, for reference. Just to keep things really fair, has anyone here got an Ouroboros that we can borrow for two weeks? The figure for the Ouroboros quoted in the CCR Table is less than we published for just the mouthpiece of the same gas dimensions as that on the Ouroboros, same bore, with extremely floppy mushrooms and large mushroom spider, large mouthpiece opening: in our case, I did publish the calibration data with the figures. However, I do not doubt that the Boris meets EN14143:2003 WOB limits - it should do comfortably. I will also publish the Meg and Inspo WOB at the same time as the Boris figures just as an extra incentive for someone with a Boris not in use every week just now. I will publish the curve for the WOB under the different loop volumes, as well as the Lissajous etc. I should add that it is possible for an entire rebreather to have a lower WOB than just the mouthpiece, but the Boris design does not seem to have the feature needed for that. Anyhow, if we test one and publish the results, then we will have numbers that are in the open and directly comparable. Alex Last edited by AD_ward9 : 7th April 2008 at 01:52. Reason: Added the Boris offer |
| (Offline) | |
| | #146 (permalink) |
| Diveshop of Horrors ![]() Current Rebreather/s: Sport Kiss MK 15.X rEvo Other CCR Azimuth Home Build Other Rebreather/s: Classic Kiss rEvo Other CCR Azimuth Home Build Join Date: Jun 2006 Location: Narragansett, Rhode Island and Hackettstown, New Jersey
Posts: 2,676
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: ISC posts test data Alex, How about the Mark-15 (with some reasonable DSV substituted for the Scott) and the CIS Mk-5P? Dave
__________________ "Silent Diving with No Bubbles and No Politics".... www.nobubblediving.com |
| (Offline) | |
| | #147 (permalink) |
| So much more to learn ![]() Current Rebreather/s: | Re: ISC posts test data Alex, Mk-16 figures are fairly well banded about, and as they come from NEDU, are not disputed. The CIS-5P would provide interesting reference points, so someone has one in Europe or I can borrow it next time I am in the USA, I promise to send it back in one piece and with the data.How about the Mark-15 (with some reasonable DSV substituted for the Scott) and the CIS Mk-5P? Dave Mk-15 is wee a bit old : it became obsolete long before EN14143 came about. Like testing an IDA 60 ![]() Alex Last edited by AD_ward9 : 7th April 2008 at 01:51. |
| (Offline) | |
| | #148 (permalink) |
| Custom Title Allowed! ![]() Current Rebreather/s: MK 15.X Ouroboros Other CCR Home Build Other Rebreather/s: Inspiration Classic Other CCR Home Build Join Date: Feb 2005
Posts: 3,328
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: ISC posts test data Alex, How about the Mark-15 (with some reasonable DSV substituted for the Scott) and the CIS Mk-5P? Dave mk15 and mk15.5 wob test results Rubicon Research Repository: Item 123456789/3525 @45m and 75 RMV Mk15 w/standard Hoses and scott mouthpiece 5.3J/L Mk15 w/redar(cooper style) Hoses and scott mouthpiece 4.1J/L Mk15 w/AGA ACSC mouthpiece and hoses 2.44J/L MK15 w/rexnord mouthpiece and mk16 hoses 2.29J/L MK15.5 w/Rexnord mouthpiece and MK16 hoses 1.97J/L @ 0.5% co2, 29 deg F Mk15 211mins MK15.5 346 mins So what does this tell us? This is what happens when experts in life support system design with a realistic budget and resources are used to design a life support system ![]() This 20year plus design out performs all new units other than boris for wob almost 6 hours scrubber life test results in 4deg C water - lol! what can the others do? lol! and a WOB as good as the best currently offered (over 20 years later) and people wonder why I still have my mk15.5! ![]()
__________________ Cave diving is a sport Wreck diving is a sport Diving in general is a sport 'Rebreather diving' is not a sport its the delusional obsession with a highly dangerous and often inappropriate piece of equipment Last edited by Drmike : 7th April 2008 at 12:49. Reason: mk15.5 wob figure updated and prism added |
| (Offline) | |
| | #149 (permalink) |
| Who loves ya, baby ![]() ![]() Current Rebreather/s: | Re: ISC posts test date I should add that it is possible for an entire rebreather to have a lower WOB than just the mouthpiece Like the Meg? ![]() Quote: The CIS-5P would provide interesting reference points, so someone has one in Europe or I can borrow it next time I am in the USA, I promise to send it back in one piece and with the data. There should be one in the UK ... ![]() NEDU also tested the MK-5p (and the Inspo) for NOAA, but results ended up in some obscure memorandum rather than a proper report. (that was the testing that Li battery warning originated from). Both WOB and duration were in that bloody memorandum, though.Quote: Mk-15 is wee a bit old That's like a dagger through his heart ... : it became obsolete long before EN14143 came about. Like testing an IDA 60 ![]() ![]() Mike and RRR, thanks for the link and filing the report, respectively.
__________________ Cheers Stefan "Political Correctness is a doctrine, fostered by a delusional, illogical minority, and rabidly promoted by an unscrupulous mainstream media, which holds forth the proposition that it is entirely possible to pick up a turd by the clean end.!" |
| (Offline) | |
| | #150 (permalink) |
| Custom Title Allowed! ![]() Current Rebreather/s: MK 15.X Ouroboros Other CCR Home Build Other Rebreather/s: Inspiration Classic Other CCR Home Build Join Date: Feb 2005
Posts: 3,328
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: ISC posts test data Mk-15 is wee a bit old : it became obsolete long before EN14143 came about. Like testing an IDA 60 ![]() Alex old - but (the MK15 configured with rexnord hoses and dsv) has better wob than all but boris and a mk15.5 and a better scrubber life than most new units so.... ![]() Ive updated the chart accordingly ![]() All hail the mighty mk15 series! ![]()
__________________ Cave diving is a sport Wreck diving is a sport Diving in general is a sport 'Rebreather diving' is not a sport its the delusional obsession with a highly dangerous and often inappropriate piece of equipment Last edited by Drmike : 7th April 2008 at 12:52. |
| (Offline) | |