| |
![]() | |
| | #1 (permalink) |
| Custom Title Allowed! Current Rebreather/s: Megalodon Classic Kiss Other Rebreather/s: Sport Kiss Join Date: Mar 2006 Location: Virginia, USA
Posts: 346
![]() ![]() ![]() ![]() ![]() ![]() | O2 Sensor Positions and Data Sets Hello, I am begining to understand how people come to make their own rebreather. The concepts and options on how to "improve" rebreather design is facinating. It's a good thing I don't have enough time on my hands to actually start building on. Like my airplanes, I'd rather fly them than spend all the time building them. Why put all the sensors essentially in one place to tell you essentially the same thing. Granted it makes figuring out if one it bad super-simple, but it seems to me if you place individual sensors around the loop you can use software programming to tell if any are far enough off "reasonable" to isolate that sensor while using the enhanced dataset to tell you much more about what is happening. In addition, while we don't currently measure anything but O2, changes in the O2 can indirectly measure the CO2 concentrations. Assume this O2 sensor configuration: * one at the mouth input (issoalted from inhalation bag) * one at the mouth output (issolated from the exhalation bag) * One at the entry to scrubber (issolated from exhalation bag * one at the exit to the scrubber (issolated from inhalation bag) Issolation is just done through one-way mushroom valves to prevent injections and additions from changing the readings significantly. Using these sensor points you can use averaging and software logic to issolate any sensors that have gone bad. You can also use the data for the computer to better understand what's actually happening in the rebreather. Using the mouthpice in and out you can tell what the diver is using. The only meaningful difference in O2 readings should be make up in expired CO2. For the in and out at the scrubber the only difference should be the removal of CO2 which will tell you about the scrubber health. I imagine you can tell alot more if you give some thoughts to having the four different readings. So, why shouldn't this work and be better than putting all the sensors together. One downside I can see is that there are more holes and wires to various parts of the loop -- increasing some complexity and opportunities for leaks. Just keeping thinking. --Paul Last edited by PaulTG2 : 15th June 2006 at 04:44. |
| (Offline) | |
| | #2 (permalink) |
| So much more to learn ![]() Current Rebreather/s: | Re: O2 Sensor Positions and Data Sets Putting sensors around the loop would give confusing readings. You also mention use of a magic wand "Software programming to tell ..". In engineering, all magic wands work against you. ![]() The way a safety engineer solves this problem is to start with a proper study of how sensors fail. We did a 5 year study. There are only 2 fault modes, out of 14 fault modes that O2 sensors have, that causes them to read high. Both of those two modes are very rare, and both can be avoided completely by using the right sensors. There are some modes that make sensors change slowly, with tendancy to low. All other modes simply cause the sensor to read low. By "low" or "high" we refer to a reading which is lower or higher than that expected after calibration. For example, a Sensor A reads 10mv in air, and Sensor B reads 8mV. In a PPO2 of 0.42, the first sensor reads 19mV so is reading low, but Sensor B reads 16mV which is reading neutral. This means that if you have calibrated the sensors correctly, then if you have n sensors, the one that gives the correct reading, is the one that reads highest. The only time one has to do another check is if the PPO2 is reported to be falling, such as on an ascent, check that the sensor you are using with the high reading, reacts to O2 injection as fast as the others. Once you can pick out which sensor is working out of n, adding sensors in one place, can give you whatever reliability you want. Cheers Alex Last edited by AD_ward9 : 15th June 2006 at 18:15. |
| (Offline) | |
| | #3 (permalink) |
| New Member Current Rebreather/s: Not Bought Yet Other Rebreather/s: Not Bought Yet Join Date: May 2006 Location: France, Montpellier
Posts: 88
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: Scrubber Sensor In other words, what we truly need it a kind of finite state machine, which will incorporate not yet another plain stupid voting logic among O2 sensors, but rather a predictable one, taking into consideration O2 sensors fault model and current dive condition (descent/ascent/hovering). Interesting idea ... |
| (Online) | |
| | #4 (permalink) |
| designer of death Current Rebreather/s: Other CCR Other Rebreather/s: Other CCR Join Date: Mar 2006 Location: kerman,california
Posts: 372
![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: Scrubber Sensor the o2 sensor response and accuracy is insufficient to do this with our current sensors. and the detector filters in the Ir sensors are not made for the pressure. i tried the reactive sensors but it was unstable under pressure, i suspect due to the membrane. the current best we seem to have is the temp stick design. i have been thinking about trying a modified detector bench for the ir type. where by you could take a standard detector and encapsulate it inside of a tube at the end with holes drilled crosswise for air flow. the filter would have to be suspended in the epoxy with some type of light shield around the edge. and all of this would have to be cleared for a high o2 environment. it would be a fun project but have too many others going on right now. rick |
| (Offline) | |
| | #5 (permalink) |
| Who loves ya, baby ![]() ![]() Current Rebreather/s: | Re: O2 Sensor Positions and Data Sets Quote: (Originally Posted by PaulTG2) Assume this O2 sensor configuration: One other item to consider is WOB. The one way way valves are among* one at the mouth input (issoalted from inhalation bag) * one at the mouth output (issolated from the exhalation bag) * One at the entry to scrubber (issolated from exhalation bag * one at the exit to the scrubber (issolated from inhalation bag) Issolation is just done through one-way mushroom valves to prevent injections and additions from changing the readings significantly. the major design problems contributing to breathing resistance, the more you add the worse it'll get. ![]()
__________________ Cheers Stefan "It is still a good day if you are on the green side of the grass! ![]() Su amigo Roberto!" Sponsor Lou in Race For Life! |
| (Offline) | |
| | #6 (permalink) |
| MEGalomaniac ![]() ![]() Current Rebreather/s: Megalodon Other Rebreather/s: Join Date: Jan 2006 Location: Pensacola, FL USA
Posts: 219
![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: O2 Sensor Positions and Data Sets Quote: (Originally Posted by AD_ward9) <snip> Alex, The way a safety engineer solves this problem is to start with a proper study of how sensors fail. We did a 5 year study. There are only 2 fault modes, out of 14 fault modes that O2 sensors have, that causes them to read high. Both of those two modes are very rare, and both can be avoided completely by using the right sensors. <snip> I'd be interested in reading the findings of the study you mention. Is it available here on Rebreather World? If not, can you point me in the right direction? Thanks! Brian
__________________ Well lit rooms will not repel determined beer ninjas. |
| (Offline) | |
| | #7 (permalink) |
| So much more to learn ![]() Current Rebreather/s: | Re: O2 Sensor Positions and Data Sets Quote: (Originally Posted by Benthic) Alex, The purpose of the work was to find good O2 sensor vendors, identify the failure modes and come up with a scheme to overcome them. The report spends a lot of time going over failure modes which one vendor seems to have far more of than the others (Teledyne, due to design as well as production issues). Exposing all the dirty laundry of one sensor vendor in too much detail is probably not good for the industry, so we will keep the report itself confidential for now.I'd be interested in reading the findings of the study you mention. Is it available here on Rebreather World? If not, can you point me in the right direction? Thanks! Brian Our clients do have copies of the report, as do each of the sensor companies whose sensors we used for the trials. We do still have some work to do on sensor approval. We are trying to accelerate the final acceptance of Analytical Industries PSR-11-33s for the Open Revolution products: they are by far the best available, based on the information we have at this time. The following link has some extracts and pictures from the report: http://www.rebreatherworld.com/megal...ghlight=Insovt Here is an extract of the things that can be helpful on the failure modes. This lists 11 failure modes overtly, but the environmental failures include 4 modes as there are 4 things to fail, so there are 14 in total. " 1.1Failure Modes Common to all Galvanic O2 Sensors
The maximum life of a sensor is the maximum of the service life and the operating life. For example, if a sensor with a 5 year shelf life and 2 year operating life is stored for 4 years before being opened and used, then it must be discarded after one year.
The trial identified the following failure modes and rates that are specific to particular sensor designs and construction:
Hope that is helpful. Cheers, Alex Last edited by AD_ward9 : 15th June 2006 at 18:16. |
| (Offline) | |
| | #8 (permalink) |
| Custom Title Allowed! Current Rebreather/s: Inspiration Classic Other Rebreather/s: Join Date: Apr 2006 Location: Denmark
Posts: 296
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: O2 Sensor Positions and Data Sets Quote: (Originally Posted by AD_ward9) The purpose of the work was to find good O2 sensor vendors, identify the failure modes and come up with a scheme to overcome them. The report spends a lot of time going over failure modes which one vendor seems to have far more of than the others (Teledyne, due to design as well as production issues). Exposing all the dirty laundry of one sensor vendor in too much detail is probably not good for the industry, so we will keep the report itself confidential for now. Hi AlexOur clients do have copies of the report, as do each of the sensor companies whose sensors we used for the trials. We do still have some work to do on sensor approval. We are trying to accelerate the final acceptance of Analytical Industries PSR-11-33s for the Open Revolution products: they are by far the best available, based on the information we have at this time. Hope that is helpful. Cheers, Alex You have made many references to the Analytical Industries PSR-11-33s being the better sensor. I have done some searching on the internet, trying to find that for my Inspiration but have only found the AI PSR 11-39-MD. Is this the same sensor that you have tested, only in a 3 pin molex version ? I don't care about price really, I just want the superiour cell. regards Johnny |
| (Offline) | |
| | #9 (permalink) |
| Rebreather Aficionado Current Rebreather/s: Other SCR Other Rebreather/s: Other SCR Join Date: Apr 2005 Location: 137km from equator
Posts: 199
![]() ![]() ![]() ![]() ![]() ![]() ![]() | Re: Scrubber Sensor Cropped out from Aii1's site. It does'nt fix your inspo. ![]()
__________________ "Silence Is Always Beautiful" |
| (Offline) | |
| | #10 (permalink) |
| So much more to learn ![]() Current Rebreather/s: | Re: Scrubber Sensor Quote: (Originally Posted by SINUS) Cropped out from Aii1's site. It does'nt fix your inspo. You have cropped only a few from the 11-33 range. Expand the scope and you will find what you are looking for.![]() AI recommend their PSR-11-39-MDs as the R22 replacement. You should find the AI PSR-11-39s an improvement over Teledyne R22s. Cheers Alex Last edited by AD_ward9 : 16th June 2006 at 08:39. |
| (Offline) | |