Re: 2 point calibration. Why?

Thanks all.

So the procedure is really just to calibrate at 100% oxygen and then check the sensor readings in air ?

I had the impression that the electronics would fit a straight line between the readings in 21% and in 100% (or something like this). Which I find rather dangerous...