| Can anyone interpret a Vision download for me Would anyone be willing to look at an APD log file from my last dive, if i e-mail it to them.
I put 3 new cells in prior to this dive. The following occured. Cell number 3 during the dive seems to be always lagging behind cells 1+2. On calibration Cells 1+2 seem to top out at 1.01 and 1.04 respectively. However Cell 3 only gets to 0.78. This is the first dive with these new cells. During the accent cell 3 was always lagging behind cells 1+2. Also it was jumping around, as you looked at the display it would read 1.19 then suddenly jump to 1.25ish. I increased the PO2 to 1.6 to see if the Cells linearity was okay, all reached 1.6 although cell 3 read the lowest. I did NOT get any cell errors. I e-mailed the file to Nicky at APD and she suggested i swap the cells posistion in the unit, to see if the problem moves. The problem i have is i am not due to dive till i take my unit abroad on 4th August. Therefore i would like to know if the cell is okay or not. If you do want to look and think you cn work it out please reply with your e-mail and i will send the APD log file. Obviously you will require the APD softwear as well. Thanks. Tim B |